Wiki

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revisionBoth sides next revision
university:tools:m2k:scopy:test-cases:logic-analyzer_temp [16 Apr 2021 03:58] – Updated for new features Trecia Agoylouniversity:tools:m2k:scopy:test-cases:logic-analyzer_temp [16 Apr 2021 04:01] Trecia Agoylo
Line 49: Line 49:
 | ::: | 4. Set logic analyzer’s settings to sample rate: 100Msps, position: 0s. Set pattern generator DIO0’s parameters to: Pattern: clock, Frequency: 2.5MHz, Phase: 0 degrees and Duty Cycle: 50%.  | {{ :university:tools:m2k:scopy:test-cases:individual_operation-step4c.png?200 |}}| Refer to the step resource picture | | ::: | 4. Set logic analyzer’s settings to sample rate: 100Msps, position: 0s. Set pattern generator DIO0’s parameters to: Pattern: clock, Frequency: 2.5MHz, Phase: 0 degrees and Duty Cycle: 50%.  | {{ :university:tools:m2k:scopy:test-cases:individual_operation-step4c.png?200 |}}| Refer to the step resource picture |
 | ::: | 5. Move the cursors to the consecutive rising edges or consecutive falling edges| {{ :university:tools:m2k:scopy:test-cases:individual_operation-step5c.png?200 |}}|The data measured by the cursor should be close to ∆t: 400ns and 1/∆t: 2.5MHz| | ::: | 5. Move the cursors to the consecutive rising edges or consecutive falling edges| {{ :university:tools:m2k:scopy:test-cases:individual_operation-step5c.png?200 |}}|The data measured by the cursor should be close to ∆t: 400ns and 1/∆t: 2.5MHz|
-| ::: | 6. Enable Lock cursor feature and measure the next edges| |}}|The data measured by the cursor should be close to ∆t: 400ns and 1/∆t: 2.5MHz|+| ::: | 6. Enable Lock cursor feature and measure the next edges| |The data measured by the cursor should be close to ∆t: 400ns and 1/∆t: 2.5MHz|
 | ::: | 7. Set logic analyzer’s settings to sample rate: 20ksps. Set pattern generator DIO0’s parameters to: Pattern: clock, Frequency: 100Hz, Phase: 0 degrees and Duty Cycle: 70%.  | {{ :university:tools:m2k:scopy:test-cases:individual_operation-step7.0c.png?200 |}}| Refer to the step resource. | | ::: | 7. Set logic analyzer’s settings to sample rate: 20ksps. Set pattern generator DIO0’s parameters to: Pattern: clock, Frequency: 100Hz, Phase: 0 degrees and Duty Cycle: 70%.  | {{ :university:tools:m2k:scopy:test-cases:individual_operation-step7.0c.png?200 |}}| Refer to the step resource. |
 | ::: | 8. Move the cursors to the rising and falling edge of the upper limit| {{ :university:tools:m2k:scopy:test-cases:individual_operation-step8c.png?200 |}}|The data measured by the cursor should be close to ∆t: 7ms| | ::: | 8. Move the cursors to the rising and falling edge of the upper limit| {{ :university:tools:m2k:scopy:test-cases:individual_operation-step8c.png?200 |}}|The data measured by the cursor should be close to ∆t: 7ms|
Line 61: Line 61:
 ^ Description                           ^ Test Steps     ^ Steps Resources ^ Expected Results ^   ^ Description                           ^ Test Steps     ^ Steps Resources ^ Expected Results ^  
  
-| Parallel Mode | 1. In logic analyzer, add a parallel decoder. Set Clock line to DIO8 and set Data lines 0-7 to DIO0 to DIO7. In pattern generator, group DIO0 to DIO7 set as number pattern. Enbale DIO8 and set to Clock with 500Hz frequency. | {{ :university:tools:m2k:scopy:test-cases:group_operation-step1.0a.png?200 |}}{{ :university:tools:m2k:scopy:test-cases:group_operation-step1.1a.png?200 |}}| The interface should look like in the "Step Resources" picture (left side).|+| Parallel Mode | 1. In logic analyzer, add a parallel decoder. Set Clock line to DIO8 and set Data lines 0-7 to DIO0 to DIO7. In pattern generator, group DIO0 to DIO7 set as number pattern. Enable DIO8 and set to Clock with 500Hz frequency. | {{ :university:tools:m2k:scopy:test-cases:group_operation-step1.0a.png?200 |}}{{ :university:tools:m2k:scopy:test-cases:group_operation-step1.1a.png?200 |}}| The interface should look like in the "Step Resources" picture (left side).|
 | ::: | 2. Run the Pattern Generator and Logic analyzer instrument. Set in the pattern generator’s instrument the desired decimal value.| {{ :university:tools:m2k:scopy:test-cases:group_operation-step2a.png?200 |}}| On the logic analyzer plot, the reading is in hex format. For reference, 50 decimal = 32 hex, 100 decimal =  64 hex and 250 decimal = FA | | ::: | 2. Run the Pattern Generator and Logic analyzer instrument. Set in the pattern generator’s instrument the desired decimal value.| {{ :university:tools:m2k:scopy:test-cases:group_operation-step2a.png?200 |}}| On the logic analyzer plot, the reading is in hex format. For reference, 50 decimal = 32 hex, 100 decimal =  64 hex and 250 decimal = FA |
 | ::: | 3. In logic analyzer, add another parallel decoder. Set Clock line to DIO8 and set Data lines 0-6 to DIO9 to DIO15.In pattern generator, group DIO9 to DIO15 set as number pattern| {{ :university:tools:m2k:scopy:test-cases:group_operation-step3.0a.png?200 |}}{{ :university:tools:m2k:scopy:test-cases:group_operation-step3.1a.png?200 |}}| The interface should look like in the "Step Resources" picture (left side).| | ::: | 3. In logic analyzer, add another parallel decoder. Set Clock line to DIO8 and set Data lines 0-6 to DIO9 to DIO15.In pattern generator, group DIO9 to DIO15 set as number pattern| {{ :university:tools:m2k:scopy:test-cases:group_operation-step3.0a.png?200 |}}{{ :university:tools:m2k:scopy:test-cases:group_operation-step3.1a.png?200 |}}| The interface should look like in the "Step Resources" picture (left side).|
university/tools/m2k/scopy/test-cases/logic-analyzer_temp.txt · Last modified: 30 Sep 2022 09:40 by Andrei Popa