 ====== Turbidity Measurement Demo (with EVAL-CN0409-ARDZ) ====== ====== Turbidity Measurement Demo (with EVAL-CN0409-ARDZ) ======
-The **EVAL-CN0409-ARDZ** shield is a low to high level water turbidity measurement system in combination with the EVAL-**ADICUP360**. It uses the **ADPD105's** ambient light rejection feature to make it ideal for applications where accurate, robust and non-contact turbidity measurements are critical.+===== General description ===== 
 +The **EVAL-CN0409-ARDZ** shield is a low to high level water turbidity measurement system in combination with the **EVAL-ADICUP360**. It uses the **ADPD105's** ambient light rejection feature to make it ideal for applications where accurate, robust and non-contact turbidity measurements are critical.
 {{ :resources:eval:user-guides:eval-adicup360:hardware:cn0409.jpg?400 |}} {{ :resources:eval:user-guides:eval-adicup360:hardware:cn0409.jpg?400 |}}
- 
-===== General description ===== 
-The **ADuCM360_demo_cn0409** 
  
 ===== Setting up the hardware ===== ===== Setting up the hardware =====
Line 16: Line 14:
  
 The user has the option to modify the solutions that are used for calibrating the device.  The user has the option to modify the solutions that are used for calibrating the device. 
-{{ :resources:eval:user-guides:eval-adicup360:reference_designs:cn0409.png?500 |}} Any values can be used for the points that are used for calibration, but a proper distribution along the 0 - 1000 NTU range must be taken into account. This values can be edited in CN0409.h file inside the include folder within the project structure.+{{ :resources:eval:user-guides:eval-adicup360:reference_designs:cn0409_default_calibration_values.png |}} Any values can be used for the points that are used for calibration, but a proper distribution along the 0 - 1000 NTU range must be taken into account. 
 +<WRAP center round info 60%> 
 +These values can be edited in **CN0409.h** file inside the include folder within the project structure
 +</WRAP> 
 +After the calibration sequence is done at least once, the calibration coefficients are saved in the controller flash memory and will be used the next time when a calibration is not done. The calibration can be repeated as desired. If the program is run for the first time and a calibration routine is not done, the program will prompt the user to manually input calibration coefficients.
        
 ===== Demo Requirements ===== ===== Demo Requirements =====
Line 50: Line 52:
   - If possible, use one properly indexed test vial.   - If possible, use one properly indexed test vial.
 <note>It is recommended to use [[https://www.coleparmer.com/i/lamotte-0290-6-turbidity-sample-test-tubes-vials-6-pk/0556366|LaMotte Test Vial]] as the size fits into the on board vial holder  </note> <note>It is recommended to use [[https://www.coleparmer.com/i/lamotte-0290-6-turbidity-sample-test-tubes-vials-6-pk/0556366|LaMotte Test Vial]] as the size fits into the on board vial holder  </note>
 +
 +===== Serial Terminal Setup =====
 +
 +{{page>/wiki/common#serial_terminal_setup&nofooter&noheader}}
  
 ===== Measurement Procedure =====      ===== Measurement Procedure =====     
resources/eval/user-guides/eval-adicup360/reference_designs/demo_cn0409.txt · Last modified: 14 Mar 2021 10:09 by Zuedmar Arceo