 +====VIBRATION EVALUATION PROGRAM WIKI GUIDE=====
  
 +=====PRODUCT SUPPORT=====
 +
 +The [[http://www.analog.com/en/evaluation/EVAL-ADIS/eb.html#SOFTWARE_AND_TOOLS|Vibration Evaluation Program]] supports PC-based evaluation of the following products with the [[adi>EVAL-ADIS|EVAL-ADIS]] Evaluation System.  
 +
 +^IMU Evaluation/EVAL-ADIS Product Support List  ^
 +|[[adi>ADIS16000|ADIS16000AMLZ]]/[[adi>ADIS16229|ADIS16229AMLZ]] |
 +|[[adi>ADIS16227|ADIS16227/PCBZ]] |
 +|[[adi>ADIS16228|ADIS16228/PCBZ]] |
 +| |
 +
 +NOTE: The [[adi>ADIS16000|ADIS16000AMLZ]] manages wireless communication with the [[adi>ADIS16229|ADIS16229AMLZ]]. 
 +
 +NOTE: The [[adi>ADIS16000|ADIS16000AMLZ]] requires an interface board to connect to the [[adi>eval-adis|EVAL-ADIS]] evaluation system. This board will be released shortly as the ADIS16COM1/PCBZ. 
 +
 +=====PC SYSTEM REQUIREMENTS=====
 +Windows XP, Vista, 7
 +
 +.NET Framework 3.5
 +
 +NOTE: Newer versions of the .NET framework do not currently support the [[/resources/eval/user-guides/inertial-mems/imu/vibrationevaluationprogram|Vibration Evaluation Program]].
 +
 +=====VIBRATION EVALUATION PROGRAM DOWNLOAD=====
 +
 +[[http://www.analog.com/en/evaluation/EVAL-ADIS/eb.html#SOFTWARE_AND_TOOLS|Vibration Evaluation Software]] installation only requires two simple step:
 +
 +1. Download, unzip and install the EVAL-ADIS USB Driver [[http://www.analog.com/en/evaluation/EVAL-ADIS/eb.html#SOFTWARE_AND_TOOLS|(click here to see the download link)]]
 +
 +2. Download, unzip and copy the Vibration*.exe file to the desired location on the test PC [[http://www.analog.com/en/evaluation/EVAL-ADIS/eb.html#SOFTWARE_AND_TOOLS|(click here to see the download link)]]
 +=====USB DRIVER INSTALLATION=====
 +After downloading the EVAL-ADIS USB Driver file, extract the SDPDrivers.exe file from the zip file, into a convenient location and then double click on it to start the process. 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:USBDriver-Install-000.png?500|}}
 +
 +When the setup wizard opens, click on **Next** to start the installation process
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:USBDriver-Install-001.png?500|}}
 +
 +Click on **Next** to accept the default driver location.
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:USBDriver-Install-002.png?500|}}
 +
 +The process will involve at least two progress bars that look like the following:
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:USBDriver-Install-003.png?500|}}
 +
 +If you encounter this type of message during the process, click on **Install** to continue with the installation.
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:USBDriver-Install-003a.png?500|}}
 +
 +Once the installation has completed, click on **Finish** to complete the process. 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:USBDriver-Install-004.png?500|}}
 +
 +=====DOWNLOAD SOFTWARE=====
 +Download to the **Vibration Evaluation Program** package from the [[adi>EVAL-ADIS|EVAL-ADIS]] home page, under SOFTWARE AND TOOLS ([[adi>en/evaluation/EVAL-ADIS/eb.html#SOFTWARE_AND_TOOLS|Click here to access]]). Extract the *.exe file from the zip file and copy it to any directory on the PC.  
 +
 +=====LAUNCH SOFTWARE=====
 +Plug the EVAL-ADIS (with ADIS16xxx/PCBZ device installed) and wait for both LED1 and LED2 to illuminate on the EVAL-ADIS.  After LED1 and LED2 illuminate, double click on the *.exe file to launch the **Vibration Evaluation Program** and access the **Main Menu**.  
 +
 +You might see the appearance of a Window that indicates a search for an SDP device (USB), but this will disappear quickly. If this error message appears, click **OK** to proceed to the **Main Screen**.
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:227-select-device.png?400|}}
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:227-main.png?600|}}
 +
 +=====PROGRAM FEATURES=====
 +====Device Selection====
 +Click on **Device**, located on the left hand side of the Menu bar, at the top of the **Main Screen**, to select the appropriate DUT (Device Under Test). In this example, the [[adi>adis16228|ADIS16228]] is in use. Note that some of the Menu bar options appear in a lighter gray color, to indicate that they are not associated with a particular device.  For example, all of the **Network** options presently only apply to the [[adi>adis16299|ADIS16229]] at this time, so they will appear in the lighter gray color when either [[adi>adsi16227|ADIS16227]] or [[adi>adsi16228|ADIS16228]] options are in use.  
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:227-main-device.png?600|}}
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:228-main-select.png?600|}}
 +
 +====Register Access====
 +The **Register Access** option on the **Menu Bar** provides direct read and write access to all of the user-accessible registers in DUT. 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:228-reg-sel.png?600|}}
 +
 +===Register Access Window Appearance===
 +
 +Here is an example of the **Register Access** window that illustrates the [[adi>adsi16228|ADIS16228]].
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_registeraccesswindow_01.png?600|}}
 +
 +===Reading Register Contents===
 +
 +Select a specific register to read the contents.
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_registeraccesswindow_regselect_01.png?600|}}
 +
 +===Writing Data to Registers===
 +
 +Use the following two steps to write a value to the register.
 +1. Enter the hex code for the register. 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_registeraccesswindow_regselect_02.png?600|}}
 +
 +2. Click on **Write**
 +
 +NOTE: Clicking on **Write** causes the Vibration Evaluation Program to write to both upper and lower bytes
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_registeraccesswindow_regselect_03.png?600|}}
 +
 +
 +
 +
 +===Single-Command Options===
 +The vibration sensor products often come with registers that support "Global Commands."  While these commands are associated with specific bits, located in user-accessible registers, the right side of the **Register Access** Window provides access to them through a drop down menu that offers each related register and a series of **Write** buttons. 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_registeraccesswindow_singlecommand_01.png?600|}}
 +
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_registeraccesswindow_singlecommand_02.png?600|}}
 +
 +
 +
 +
 +
 +====Network====
 +
 +Coming soon.....
 +====Alarms====
 +The **Alarm > Alarm Settings** option on the **Menu Bar** provides a more convenient method for configuring the Spectral Alarm functions.  This provides a more convenient method for tuning this function, in comparison with the single-register access method associated with **Register Access** window. 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:227-alrm.png?600|}}
 +
 +Selecting **Alarms > Alarm Settings** will cause the following window to open:
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_alarms_settings_01.png?600|}}
 +
 +Select boxes in the matrix and enter values that are associated with the magnitude of the output data and FFT bin numbers. 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_alarms_settings_02.png?600|}}
 +
 +Click on **Write to DUT** to update all registers that are associated with these entries.
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_alarms_settings_03.png?600|}}
 +
 +Close this window, then go back in. The values will not appear automatically. Click on **Read from DUT** to make sure the settings are still in place.
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_alarms_settings_04.png?600|}}
 +
 +The **Alarms > Alarm Status Form** option provides a convenient monitor for the conditions. The dashes will change to green (no alarm), yellow ("warning" alarm, associated with Level 1) or red ("critical" alarm, associated with Level 2), depending on the conditions, after a data capture event completes. 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_alarms_statusform_01.png?400|}}
 +
 +====Data Capture====
 +The **Data Capture** window provides user inputs for file location, base file name and for resetting the file count.  
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_datacapturewindow_01.png?500|}}
 +
 +When this function is active (See the [[/resources/eval/user-guides/inertial-mems/imu/vibrationevaluationprogram#enable_data_capture|Enable Data Capture checkbox]], located in the **Main Screen**), each trigger will cause the creation of a new file that contains the FFT result and FFT Header information. Notice the increments in the file count, after three clicks on the **Start** button. 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:227-main-datalog-file.png?600|}}
 +
 +This counter is also in the **Data Capture Window**
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_datacapturewindow_02.png?500|}}
 +====Tools====
 +The **Tools** option in the **Menu Bar** offers two options: **USB** and **SPI**. 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:227-tools-menu.png?600|}}
 +
 +Use the **USB** option to manually connect or disconnect to the USB port on the [[adi>eval-adis|EVAL-ADIS]]. 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_tools_usb_01.png?400|}}
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_tools_usb_02.png?400|}}
 +
 +Use the **SPI** option to adjust the timing in between the [[adi>EVAL-ADIS|EVAL-ADIS]] and the DUT. this should not be required for normal operation but the "typical" settings are offered in the following picture: 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:227-spi-utilities.png?400|}}
 +
 +
 +
 +
 +
 +
 +
 +
 +
 +
 +====Demo====
 +
 +Visit the [[resources/eval/user-guides/inertial-mems/imu/vibrationdemo|ADIS16229 Vibration Demo Wiki Guide]] for more details on this function. 
 +====About====
 +This option offers the revision and some codes that might be useful when seeking technical support. 
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:227-version-1-2.png?400|}}
 +
 +====Data Collection Mode====
 +The [[adi>adsi16227|ADIS16227]], [[adi>adsi16228|ADIS16228]] and [[adi>adsi16229|ADIS16229]] all have four basic modes of data collection: Manual FFT, Automatic FFT, Manual Time (Time Domain) and Real-Time. Inside of these products, these modes are typically related to the settings in the REC_CTRL1 register.  The **Main Screen** offers a drop-down selection menu for these modes, along with a **Start** button to begin operation.   
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:vep_wiki_mainscreen_modeselection_01b.png?600|}}
 +
 +====Waveform Display====
 +The **Waveform Display** enables quick access to sensor data.
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:227-wavfrms.png?600|}}  
 +
 +
 +====Enable Data Capture====
 +Checking the box next to **Enable Data Capture** results in every capture of data being copied into a data file, according to the configuration settings in the **Data Capture Window**.
 +
 +{{:resources:eval:user-guides:inertial-mems:imu:227-enable-datalog.png?600|}}
 +
 +
 +
 +
 +
 +
 +
 +=====SOFTWARE REVISION HISTORY=====
 +
 +^REVISION  ^RELEASE DATE ^SUMMARY OF UPDATES ^
 +|v1.3.0| 2/17/2014| See the [[/resources/eval/user-guides/inertial-mems/imu/vibrationevaluationprogram?&#reported_issues_solutions|Reported Issues & Solutions Table]] for  list of issues being addressed |
 +|v1.2.0| 1/25/2014| See the [[/resources/eval/user-guides/inertial-mems/imu/vibrationevaluationprogram?&#reported_issues_solutions|Reported Issues & Solutions Table]] for  list of issues being addressed |
 +|v1.1.9| 12/22/2013| See the [[/resources/eval/user-guides/inertial-mems/imu/vibrationevaluationprogram?&#reported_issues_solutions|Reported Issues & Solutions Table]] for  list of issues being addressed |
 +|v1.1.8| N/A | Internal test version, never published online |
 +|v1.1.7| 12/10/2013| Added partial support for the ADIS16227 (Manual FFT mode only) |
 +|v1.1.6| 11/1/2013 | Extended the wait time to 4 seconds, for each sensor in the Network Scan |
 +|v1.1.5| N/A | This version corrects a sizing problem in the main waveforms, along with a couple of bugs in other windows. |
 +|v1.1.4| N/A | Updates for internal use only |
 +|v1.1.3| N/A | Added delays to Periodic Mode Exit routine register writes. |
 +|v1.1.2| N/A| Added a dialog box for a user to specify the Update Interval of the Network Periodic Mode. |
 +|v1.1.1| 10/11/2013| Corrected the GUI to indicate "not busy" after a communication is canceled by the user. |
 +| | | Added exception handling routines to prevent a program crash and to show the Reconnect Dialog if the USB is disconnected. |
 +| | | Expanded the vertical axis range options on the FFT plots (Main Menu) to go down to 0.0001g |
 +|v1.1.0| N/A| Added "Scan All Channels" option to find nodes that are on different frequency channels |
 +| | | Expanded the vertical axis range options on the FFT plots (Main Menu) to go down to 0.001g |
 +|v1.0.3| 9/10/2013 | Set the plot scale to +-2g for the first switch to time domain data. This ensures that data is visible on the plot. Subsequent plot mode changes set the scale to the previous user selected value for a particular mode.|
 +| | | Set the Mode combo-box to ‘Periodic’ or ‘Manual’ when a demonstration mode is selected. This ensures that the plot mode is appropriate for the data displayed during a demonstration loop.|
 +|v1.0.2 | 8/21/2013| Add sensor node selection tabs to Register Access and Alarm menus|
 +| | | NOTE: First version posted online at www.analog.com/EVAL-ADIS|
 +|v1.0.1 | 8/15/2013| Updated register names to match the [[adi>adis16229|ADIS16229]] datasheet|
 +| | | Increased period mode sleep cycle time to 10 seconds|
 +|v1.0.0 | 8/1/2013| Initial Release|
 +
 +=====REPORTED ISSUES, UPGRADE REQUESTS & SOLUTIONS=====
 +^REVISION  ^ # ^DATE ^STATUS ^DEVICE ^ DESCRIPTION & RELATED NOTES ^SOLUTION ^
 +| | | | | | | | 
 +|v1.3.1 | 01 | 7/29/14 |CONFIRMED | [[adi>ADIS16228|ADIS16228]] | Real-time mode does not work after executing an **Auto-null** command in the **Register Access** menu. {{:resources:eval:user-guides:inertial-mems:imu:vep-v1_3_1-problemreport_adis16228.pdf|Click here for more details}} | After executing an **Auto-null** command, select Manual FFT mode first, then **Real-time** mode. | 
 +|v1.3.0 | 01 | 2/24/14| CONFIRMED | [[adi>ADIS16228|ADIS16228]] | Real-time only supports data collection at a rate of 1.26kSPS | v1.4.0 will address this. For now, use AVG_CNT settings that are __>__ 4 (Sample rate = 1260 SPS) | 
 +|v1.2.0 | 04 | 2/12/14 | CLOSED | [[adi>ADIS16000|ADIS16000]] | Connection with [[adi>ADIS16229|ADIS16229]] is hard to establish | SPI Timing violation, click here for more details |
 +|v1.2.0 | 03 | 1/27/14 | CONFIRMED | [[adi>ADIS16228|ADIS16228]] | **Time Domain** mode in this package seems to translate into **Manual Time Capture** in the [[adi>ADIS16228|ADIS16228]] datasheet. Can the software be consistent with the datasheet? | v1.3.0 (or later), Not released yet |
 +|v1.2.0  | 02 | 1/27/14 | CONFIRMED | [[adi>ADIS16228|ADIS16228]] | The **Data Capture** function does not appear to work when using **Time Domain** mode | v1.3.0 (or later), Not released yet |
 +|v1.2.0  | 01 | 1/27/14 | CONFIRMED | [[adi>ADIS16228|ADIS16228]] | When using the **Data Capture** function, when in **Real-Time** mode, can this produce one continuous file for the data samples? | v1.3.0 (or later), Not released yet |
 +|v1.1.9 | 01 | 1/9/14 | CLOSED | [[adi>ADIS16228|ADIS16228]] | Add all of the FFT Header registers to the Data Capture file | v1.2.0 (or later) | 
 +|v1.1.7 | 06 | 12/17/13 | CLOSED | [[adi>ADIS16228|ADIS16228]] | There appear to be a few typos in the register names in **Register Access** menu, as they do not match the names given in the [[adi>ADIS16228|ADIS16228]] datasheet | v1.1.9 (or later) |
 +|v1.1.7 | 05 | 12/17/13 | CLOSED | [[adi>ADIS16228|ADIS16228]] | Exiting **Periodic FFT** mode causes the program to jump into the **Main Screen > Tools > USB Menu**. When closing that Window, it engages in a 10 second countdown while allowing no user input.  | v1.1.9 (or later). |
 +|v1.1.7 | 04 | 12/17/13 | CLOSED | [[adi>ADIS16228|ADIS16228]] | Changes in the Mode Selection drop-down menu, located in the **Main Screen**, do not seem to track the related register settings in the **Register Access** menu  | v1.1.9 (or later). |
 +|v1.1.7 | 03 | 12/13/13 | CLOSED | [[adi>ADIS16228|ADIS16228]]  | **Manual FFT** only seems to support 20480 SPS sample rates. For more details, click on the following file: {{:resources:eval:user-guides:inertial-mems:imu:vep-problemreport-v1-1-7-adis16228-manualfftmultrecord.pdf|Problem Report Details}} | v1.1.9 (or later). | 
 +|v1.1.7 | 02 | 12/10/13  | CLOSED | [[adi>ADIS16228|ADIS16228]] | The **Alarm Status Form** does not correctly display alarm status.  Click on this file for more details: {{:resources:eval:user-guides:inertial-mems:imu:vep-problemreport-v1-1-7-adis16228-alarmstatusform.pdf| Problem Report Details}} | No issue found. Open this file for more details: {{:resources:eval:user-guides:inertial-mems:imu:vep-problemresolutionreport-v1-1-7-adis16228-alarmstatusform.pdf| Problem Resolution Details}} |
 +|v1.1.7 | 01 | 12/10/13  | CLOSED | [[adi>ADIS16227|ADIS16227]] | **Periodic FFT** appears to lock the software up and does not appear to produce FFT results | v1.2.0 (or later) |
 +|v1.1.6 | 06 | 12/9/13  | CONFIRMED | [[adi>ADIS16000|ADIS16000]] | There appear to be a few typos in the register names in **Register Access** menu, as they do not match the names given in the [[adi>ADIS16000|ADIS16000]] datasheet | v1.1.10, release date estimate = 1/10/2014 |
 +|v1.1.6 | 05 | 12/9/13  | CLOSED | [[adi>ADIS16000|ADIS16000]] | Not able to add [[adi>adis16229|ADIS16229]] devices to pages 2-6 in the [[adi>ADIS16000|ADIS16000]] | No issue found, but this will be closely scrutinized in the next two versions: v1.1.10 and v1.2.0 |
 +|  |  |  |  |  | ADI: This function was verified in earlier versions of the software. Re-test of this function is in the test queue. |  | 
 +|v1.1.6 | 04 | 12/9/13  | CONFIRMED | [[adi>ADIS16000|ADIS16000]] | There are a number of cases where the software becomes non-responsive, with particular sensitivity observed when using Periodic FFT. | v1.2.0, release date estimate = 1/24/2014. For present revision, use **Manual FFT** mode only. |
 +|  |  |  |  |  | ADI: Investigation has revealed that the present approach for simultaneous sensor monitoring and user input management is not stable. The solution to this will require substantial code changes, which are in progress. |  |  
 +|v1.1.6 | 03 | 12/9/13  | CLOSED| [[adi>ADIS16000|ADIS16000]] | Sample rate settings in the **AVG_CNT** register do not seem to cause updates on the horizontal axis, of the x-axis data. | Use version v1.1.9 or later. |
 +|v1.1.6 | 02 | 12/9/13  | CONFIRMED | [[adi>ADIS16000|ADIS16000]] | Real-time data logging is not working | This was not part of the original plan for this package, but is under consideration for v1.4.0 or later. |
 +|  |  |  |  |  |  |  | 
 +|v1.1.6 | 01 | 12/9/13  | CLOSED | [[adi>ADIS16000|ADIS16000]] | Waveforms updates do not always respond to **Start** button presses, in any mode. | Use v1.2.0 or later |
 +|  |  |  |  |  |  |  | 
 +|  |  |  |  |  | **__STATUS CODE DEFINITIONS__** |  | 
 +|  |  |  |  |  | **OPEN** means that the issue observation has been received, but has not been independently confirmed by ADI |
 +|  |  |  |  |  |**CONFIRMED** means that the issue has been independently confirmed and ADI is working on a solution path.|
 +| | |  | | | **CLOSED** means that the issue has been resolved|
 +|  |  |  |  |  |  | 
 +
 +
 +
 + 
resources/eval/user-guides/inertial-mems/imu/vibrationevaluationprogram.txt · Last modified: 17 Feb 2021 05:58 by Juan Chong